The DMC-1410 Calibrator fully documents any calibration easily while performing the normal job. No special software is required! It features the ability to test for pass/fail criteria, measure and simulate thermocouples, RTDs, current, voltage, and frequency, and source pulse trains.
In addition, the built-in 250 Ohm resistor for HART compatibility with smart transmitters and PLCs, full fuseless protection, and a serial communications port for full control with ASCII commands.
- Stores calibration data for 50 instruments in non-volatile memory
- Print calibration certificates without a PC
- ±0.4°C thermocouple/±0.3°C RTD accuracy
- ±0.015% of reading electrical accuracy Thirteen (13) T/C types and Thirteen (13) RTD types, plus YSI 400
- Simultaneous read and source thanks to isolated mA/V read-back circuit for complete transmitter calibration
- Full Function 4 to 20 mA read, source, simulate and read/power modes with ±0.015% of reading accuracy
- Source or read 0 to 20 volts (±0.015% of reading accuracy)
- Direct keyboard entry enables quick output changes
- All source modes can be programmed with dedicated setpoints
- Built-in 250 Ohm resistor for HART compatibility with smart transmitters and PLCs
- Direct entry of custom RTD coefficients(R0,A,B,C)
- RTD function compatible with all pulsed transmitters
- RS232 serial interface
- Supplied in full rubber boot; optional carrying case available
- Includes NIST Certificate with calibration data
- Communication port compatible with both Fluke 700 Series and Beta Calibrator pressure modules.
Got a question or want a quote? Call or use our contact us form